This is the only Random Access Automatic Analyzer in its price range provided with a Continuous Access feature that permits the operator to insert a STAT ample anywhere on the tray at any time. Operates in Batch Mode (test by test) or in Random Mode (patient by patient) as required. The reagent tray is cooled and the sampling probe is provided with an inductive level detector for reagents and samples (with primary tubes only). Optional adapters permit the use of 2.5 mL tubes for the reagents. The instrument is supplied with clinical chemistry software as standard - different software versions are available for immunochemistry or other tests. |
|
| Access Mode | Patient by patient, full random access, STAT analysis |
| Operation Mode | By profile, test by test (optional), walk away |
| Analysis Type | End point, initial rate, kinetic, bichromatic, differential (sample blank), turbidimetric, multi-standard - up to two reagents per test |
| Patients File | Up to 24 tests/patient, up to 4000 reports on 1.44 Mb floppy disk |
| STAT Time | Immediate |
| Reports | Alphanumeric, patient by patient, or test by test |
| Sampling System | SIP and DIP with automatic probe washing device between each sample and each test |
| Automatic Dilution | Samples can be diluted and repeated automatically |
| Predilution | User selectable dilution ratio, automatic (only for immunochemistry) |
| Optional | External computer, DAU software, immunochemistry software, reagent tubes adapter |
| Sample Tray | Selectable, 20 primary tubes + 18 sample cups, or 58 sample cups only |
| Reagent Tray | Removable, 15 positions for 45 mL containers, cooling system by Peltier cells |
| Reactions Cells | 10 removable racks for 180 tests |
| Throughput | Up to 120 tests/hour |
| Photometer | 6 interference filters from 340 to 620 nm + 1 free position for optional filter; resolution 0.0001 Abs; accuracy better than 1% CV; 50µ flow through cell; Peltier cells temperature control |
| Reagent Volume | Variable from 5 to 700 µL |
| Sample Volume | Variable from 2 to 100 µL |
| Printer | Built in 40 column high speed thermal printer with graphic capability |
| CPU | 386 CPU with integrated VGA adapter and interface; 8 Mb disk on chip flash; parallel port and serial port |
| Dimensions | 68 cm (W) x 52 cm (D) x 35 cm (H); weight 35 kg |
| Power | 110-230V; 50/60 Hz; power consumption 400W |
Cat No. BMD-RAT-0001 |
|